Basic instrument parameters


Prime Focus Instrument (PFI)

Item Description
Field of view ∼1.38 deg (hexagonal - diameter of circumscribed circle)
Field of view area ∼1.25 deg2
Input F number to fiber 2.8
Fiber core diameter 1 127 μm (1.1 arcsec at the FoV center, ∼1.0 arcsec at the edge)
Positioner pitch 8 mm (90.4 arcsec at the FoV center, 82.4 arcsec at the edge)
Positioner patrol field 9.5 mm diameter (107.4 arcsec at the FoV center, 97.9 arcsec at the edge)
Fiber minimum separation 2 ∼30 arcsec
Fiber configuration time ∼130 s
Number of fibers 3 Science fibers 2385 (2394), Fiducial fibers 92 (96)
Fiber density ∼2000 deg-2 / ∼0.6 arcmin-2
Number of AG camera 4 6
Field of view of AG camera ∼5.1 arcmin2 per one camera
Sensitivity of AG camera gaia ∼21.7 AB mag for S/N∼5 in 5 second exposure
  1. This is a diameter of the sky projected on to the fiber core not directly but through a microlens with a magnification of 1.28. According to the plate scale without microlens, 100μm subtends ∼1.1 arcsec on the sky.
  2. The minimum separation includes a physical limitation and a margin for collision avoidance.
  3. Excluding the broken fibers. Number in parentheses is from the design.
  4. The AG cameras are mounted surrounding the hexagonal shaped array of fiber positioners.

Spectrograph System (SpS)

Blue Red (Low res.) Red (Med res.) NIR
Spectral coverage 380 - 650 nm 630 - 970 nm 710 - 885 nm 940 - 1260 nm
Dispersion ∼0.7 Å/pix ∼0.9 Å/pix ∼0.4 Å/pix ∼0.8 Å/pix
Spectral resolution ∼2.1 Å ∼2.7 Å ∼1.6 Å ∼2.4 Å
Resolving power ∼2500 (@500nm) ∼3000 (@800nm) ∼5500 (@800nm) ∼4500 (@1100nm)
Spectrograph throughput 4 ∼49% (@500nm) ∼52% (@800nm) ∼48% (@800nm) ∼39% (@1100nm)