CCD troubles and the solutions
There are some typical cases of the CCD-related troubles.
- CCD would not respond to the SOSS command (i.e., integration
would
not starts after wiping the CCD).
- Incoming image is the same as the last image taken.
- All images look strange. E.g., every row of the image is
misaligned.
Check the cause for the case 1
- Cannot issue commands relevant to CCD such as "Vcheck" and "Init"
from UI. You will find an error message immediately after issuing the
command.
- There are some errors, or "[E]" messages in CE and/or M3.
- No NI and/or FT software are running when checking on the
"process
check" button on the FALCONMENU.
Solution for the case 1
- Press "cancel" in the SOSS.
- Type "CA" and <return> in CE terminal to CAncel the CE
commands.
- Type "CA" and <return> in M3 terminal to CAncel the M3
commands.
- If some software is missing in the "process check", quit it and
restart
it from the FALCONMENU.
Check the cause for the cases of 2 and 3
Press "Vcheck" on the CCD tab screen on the UI. If you receive an
error,
the CCD power has turned off for some reasons.
Check the voltage displayed in Messia_log window. If it shows any bad
channels, the power went off.
Solution for the cases of 2 and 3
Turn off the CCD power from the APC netscape power manager, and then
powet it up after wating for a while (say, ~10 sec).
Press "Vcheck" again to see if there is no bad channel detected.